- 12 -
Negative % (Torque) will be displayed as shown in Figure 17:
Figure 17
cP (Viscosity) and SS (Shear Stress) values will be displayed as "- - - -" when the % (Torque) is
below zero.
Figure 18 is an example of the printed output of each of the above conditions.
RPM=50 MD=RV SPDL=29 %=51.4 cP=10280 D/CM2=1285 1/SEC=12.5 T=25.5C
Normal Print String
RPM=50 MD=RV SPDL=29 %=EEEE cP=EEEE D/CM2=EEEE 1/SEC=12.5 T=25.5C
Over-Range
?RPM=50 MD=RV SPDL=29 %=5.2 cP=1040 D/CM2=130 1/SEC=12.5 T=25.5C
Under-Range
RPM=50 MD=RV SPDL=29 %=-0.1 cP=---- D/CM2=---- 1/SEC=0.0 T=25.5C
Negative Torque
Figure 18
II.7 Temperature Display
The DV-II+ will display the temperature measured by its RTD temperature probe. Temperature may
be displayed in either ˚C (Centigrade) or ˚F (Fahrenheit) units depending upon the start-up
operations. If the DV-II+ is started normally, the default temperature display will be in ˚C
(Centigrade) units as shown in the Figure 19:
Figure 19
However, if you press and hold the SELECT DISPLAY key while turning on the DV-II+ the
temperature will be displayed in ˚F (Fahrenheit).
If you turn on the DV-II+ with the temperature probe disconnected, or remove the temperature probe
at any point after power-up, the display will indicate - - - -°C.
The four dashes indicate the absence of the probe. If you were displaying temperature in
Fahrenheit units the C would be displayed as an F.