Brookfield DV-II+ Programmable Viscometer Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Instruments de mesure Brookfield DV-II+ Programmable Viscometer. Brookfield DV-II+ Programmable Viscometer User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 75
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
Brookfield Engineering Labs., Inc. Page 1 Manual No. M/97-164-F1 102
BROOKFIELD DV-II+
PROGRAMMABLE VISCOMETER
Operating Instructions
Manual No. M/97-164-F1102
BROOKFIELD ENGINEERING LABORATORIES, INC.
11 Commerce Boulevard, Middleboro, MA 02346-1031 USA
SPECIALISTS IN THE
MEASUREMENT AND
CONTROL OF VISCOSITY
T
EL
508-946-6200
F
AX
508-946-6262
or 800-628-8139 (USA only)
I
NTERNET
www.brookfieldengineering.com
Vue de la page 0
1 2 3 4 5 6 ... 74 75

Résumé du contenu

Page 1 - Manual No. M/97-164-F1102

Brookfield Engineering Labs., Inc. Page 1 Manual No. M/97-164-F1 102BROOKFIELD DV-II+PROGRAMMABLE VISCOMETEROperating InstructionsManual No. M/97-

Page 2

Brookfield Engineering Labs., Inc. Page 10 Manual No. M/97-164-F1102II. GETTING STARTEDII.1 AutozeroBefore readings may be taken, the Viscometer mu

Page 3

Brookfield Engineering Labs., Inc. Page 11 Manual No. M/97-164-F1 102II.2 SELECTSPINDLESELECTSPINDLE Spindle SelectionLVDV-II+ Viscometers are prov

Page 4

Brookfield Engineering Labs., Inc. Page 12 Manual No. M/97-164-F1102 II.3 SETSPEED Speed Selection, Setting, RunningThere are 54 speeds programmed

Page 5

Brookfield Engineering Labs., Inc. Page 13 Manual No. M/97-164-F1 102If an ARROW key is pressed and held the DV-II+ will scroll up (or down) throug

Page 6 - + is in the OFF position

Brookfield Engineering Labs., Inc. Page 14 Manual No. M/97-164-F1102SR 40.0 20.1C6.0RPM % 15.6Figure 13One more press of the SELECT DISPLAY k

Page 7

Brookfield Engineering Labs., Inc. Page 15 Manual No. M/97-164-F1 102II.6 Out of RangeThe DV-II+ gives indications for out-of-range operation. Whe

Page 8

Brookfield Engineering Labs., Inc. Page 16 Manual No. M/97-164-F1102II.7 Temperature DisplayThe DV-II+ displays the temperature measured by its RTD

Page 9

Brookfield Engineering Labs., Inc. Page 17 Manual No. M/97-164-F1 102Figure 21When printing via the parallel port, please note that if a printer is

Page 10 - + default screen:

Brookfield Engineering Labs., Inc. Page 18 Manual No. M/97-164-F1102Non-Newtonian fluid behavior can result in the measured viscosity changing if t

Page 11

Brookfield Engineering Labs., Inc. Page 19 Manual No. M/97-164-F1 102III. OPTIONSIII.1 TABOPTIONS Introduction to OPTIONSThe OPTIONS/TAB key provid

Page 12 - 6.0RPM12 % 15.6

Brookfield Engineering Labs., Inc. Page 2 Manual No. M/97-164-F1102TABLE OF CONTENTS I. INTRODUCTION ...

Page 13

Brookfield Engineering Labs., Inc. Page 20 Manual No. M/97-164-F1102TABOPTIONS OPTIONS/TAB - Toggles between options.AUTORANGEENTER ENTER/AUTORANGE

Page 14 - SR 40.0 20.1C

Brookfield Engineering Labs., Inc. Page 21 Manual No. M/97-164-F1 102Selecting an OptionThe following is a quick reference for entering and using t

Page 15 - 10 RPM % 8.2

Brookfield Engineering Labs., Inc. Page 22 Manual No. M/97-164-F1102L°F(FAHRENHEIT) CGS UNITSLSEQUENTIAL PRINT SERIALLPRINT SERIAL DATA AVERAGINGL°

Page 16 - 10 RPM P% 19.7

Brookfield Engineering Labs., Inc. Page 23 Manual No. M/97-164-F1 102L°F(FAHRENHEIT) CGS UNITSFigure 28A press of the OPTIONS/TAB key at this point

Page 17 - CHECK CONNECTION

Brookfield Engineering Labs., Inc. Page 24 Manual No. M/97-164-F1102III.2.3.2 Custom SpeedsPressing the ENTER/AUTORANGE key when Custom Speeds is

Page 18

Brookfield Engineering Labs., Inc. Page 25 Manual No. M/97-164-F1 102Step 2 is repeated indefinitely until the viscometer is shut off or the user s

Page 19 - III. OPTIONS

Brookfield Engineering Labs., Inc. Page 26 Manual No. M/97-164-F1102III.3.1 Time to Stop On entry, the user is presented with the following screen

Page 20

Brookfield Engineering Labs., Inc. Page 27 Manual No. M/97-164-F1 102That done, the user is presented with the following screen: TIMED STOPENTER

Page 21 - TIME TO TORQUE

Brookfield Engineering Labs., Inc. Page 28 Manual No. M/97-164-F1102standard print string to the attached printer as many times as the user presses

Page 22 - CGS UNITS

Brookfield Engineering Labs., Inc. Page 29 Manual No. M/97-164-F1 102a screen similar to Figure 47 for the duration of the timed torque run: TORQUE

Page 23

Brookfield Engineering Labs., Inc. Page 3 Manual No. M/97-164-F1 102I. INTRODUCTIONThe Brookfield Programmable DV-II+ Viscometer measures fluid vis

Page 24 - PRINT PARALLEL

Brookfield Engineering Labs., Inc. Page 30 Manual No. M/97-164-F1102However, upon starting the timed operation, the DV-II+ will output an initial d

Page 25 - 10 RPM A% 19.7

Brookfield Engineering Labs., Inc. Page 31 Manual No. M/97-164-F1 102Activating print selections in the Print mode can only be done by exiting to t

Page 26

Brookfield Engineering Labs., Inc. Page 32 Manual No. M/97-164-F1102Refer to Section IV for information on how to create B.E.V.I.S. programs on a P

Page 27 - TIMED STOP DONE

Brookfield Engineering Labs., Inc. Page 33 Manual No. M/97-164-F1 102RUN PGM SLOT#1 PRESS ENTERFigure 60Using the UP/DOWN ARROWkeys, select one of

Page 28 - SELECT RPM: 30

Brookfield Engineering Labs., Inc. Page 34 Manual No. M/97-164-F1102CP 123.4 55.1C10 RPM PGM% 63.7Figure 65The DVLOADER Software is used to creat

Page 29 - Figure 35

Brookfield Engineering Labs., Inc. Page 35 Manual No. M/97-164-F1 102IV. DVLOADER SOFTWAREThe DVLOADER software is a WINDOWS-based program provide

Page 30 - SET MIN’S: 00

Brookfield Engineering Labs., Inc. Page 36 Manual No. M/97-164-F1102IV.2 Description of B.E.V.I.S. CommandsCommand Code Required Parameter Command

Page 31

Brookfield Engineering Labs., Inc. Page 37 Manual No. M/97-164-F1 102IV.3 Creating a B.E.V.I.S. ProgramStart the DVLOADER software by clicking on i

Page 32

Brookfield Engineering Labs., Inc. Page 38 Manual No. M/97-164-F1102Figure 67 shows the grid where the operator programs arecreated. It is used to

Page 33 - PRESS A KEY

Brookfield Engineering Labs., Inc. Page 39 Manual No. M/97-164-F1 102Click the Delete button to delete the B.E.V.I.S. command in the selected row o

Page 34 - 10 RPM PGM% 63.7

Brookfield Engineering Labs., Inc. Page 4 Manual No. M/97-164-F1102I.1 ComponentsComponent Part Number QuantityDV-II+ Viscometer varies 1Model S La

Page 35 - IV. DVLOADER SOFTWARE

Brookfield Engineering Labs., Inc. Page 40 Manual No. M/97-164-F1102Program 2: For use with an external temperature controllerCommand Command Descr

Page 36

Brookfield Engineering Labs., Inc. Page 41 Manual No. M/97-164-F1 102SSN Set viscometer speed 5.0 run at 50 RPMWTI Wait for time interval 00:10 wai

Page 37 - Figure 66

Brookfield Engineering Labs., Inc. Page 42 Manual No. M/97-164-F1102Appendix A - Cone/Plate Viscometer Set-UpThis Cone/Plate version of the DV-II+

Page 38 - Figure 67

Brookfield Engineering Labs., Inc. Page 43 Manual No. M/97-164-F1 102Bath/CirculatorBathInletBathOutletSampleCup(CPE-44YorCPE-44P)CupOutletCupInlet

Page 39 - • Set PC PROG to “OFF.”

Brookfield Engineering Labs., Inc. Page 44 Manual No. M/97-164-F1102Moves Awayfrom Hit Point(clockwise) LEFTx Moves Towards Hit Point(counter-cloc

Page 40 - (time to torque measurement)

Brookfield Engineering Labs., Inc. Page 45 Manual No. M/97-164-F1 102Cone Sample VolumePart No.CPE-40 0.5 mlCPE-41 2.0 mlCPE-42 1.0 mlCPE-51 0.5 ml

Page 41

Brookfield Engineering Labs., Inc. Page 46 Manual No. M/97-164-F1102Appendix B - Viscosity RangesViscosity ranges shown are for operational speeds

Page 42 - Figure A1

Brookfield Engineering Labs., Inc. Page 47 Manual No. M/97-164-F1 102 DIN Adapter Accessory Spiral Adapter Cone/Plate Viscometer Helipath with T-Ba

Page 43 - Figure A4

Brookfield Engineering Labs., Inc. Page 48 Manual No. M/97-164-F1102The second consideration involves the mechanics of fluid flow. All rheological

Page 44 - Figure A6

Brookfield Engineering Labs., Inc. Page 49 Manual No. M/97-164-F1 102Appendix C - Variables in Viscosity MeasurementsAs with any instrument measure

Page 45

Brookfield Engineering Labs., Inc. Page 5 Manual No. M/97-164-F1 102I.3 SpecificationsSpeeds: Interleaved: LV/RV (18 speeds)Sequential: LV/RV (18 s

Page 46 - Appendix B - Viscosity Ranges

Brookfield Engineering Labs., Inc. Page 50 Manual No. M/97-164-F1102container wall and the spindle surface.A repeatable viscosity test should contr

Page 47 - (1-100 RPM

Brookfield Engineering Labs., Inc. Page 51 Manual No. M/97-164-F1 102Appendix D - Spindle and Model CodesEach spindle has a two digit entry code wh

Page 48

Brookfield Engineering Labs., Inc. Page 52 Manual No. M/97-164-F1102SC4-14 14 125 0.4SC4-15 15 50 0.48SC4-16 16 128 0.2929SC4-18 18 3.2 1.32SC4-21

Page 49

Brookfield Engineering Labs., Inc. Page 53 Manual No. M/97-164-F1 102The full scale viscosity range for any DV-II+ model and spindle may be calcula

Page 50

Brookfield Engineering Labs., Inc. Page 54 Manual No. M/97-164-F1102Appendix E - Calibration ProceduresThe accuracy of the DV-II+ is verified using

Page 51 - Table D1

Brookfield Engineering Labs., Inc. Page 55 Manual No. M/97-164-F1 102 Brookfield Viscosity Standard Fluid General InformationWe recommend that Broo

Page 52 - Table D2

Brookfield Engineering Labs., Inc. Page 56 Manual No. M/97-164-F11024) Allow 30 minutes for the viscosity standard, sample chamber and spindle to r

Page 53

Brookfield Engineering Labs., Inc. Page 57 Manual No. M/97-164-F1 1024) The viscosity standard fluid, together with the spindle, should be immersed

Page 54 - °C, 93.3°C, 149°C

Brookfield Engineering Labs., Inc. Page 58 Manual No. M/97-164-F1102 Interpretation of Calibration Test Results:When verifying the calibration of t

Page 55

Brookfield Engineering Labs., Inc. Page 59 Manual No. M/97-164-F1 102The calibration of the Brookfield Viscometer/Rheometer is determined using a 6

Page 56 - + Viscometer

Brookfield Engineering Labs., Inc. Page 6 Manual No. M/97-164-F1102I.4 InstallationNote: “IQ, OQ, PQ”, a guideline document for installation, opera

Page 57

Brookfield Engineering Labs., Inc. Page 60 Manual No. M/97-164-F1102many viscometer users the true viscosity is not as important as a repeatable da

Page 58 - SMC = 10 from Table D1

Brookfield Engineering Labs., Inc. Page 61 Manual No. M/97-164-F1 102Appendix F - Speed SetsThe following speed sets and custom speeds are selectab

Page 59 - LV Guardleg

Brookfield Engineering Labs., Inc. Page 62 Manual No. M/97-164-F1102Appendix G - CommunicationsWhen using the Brookfield Computer Cable (Brookfield

Page 60

Brookfield Engineering Labs., Inc. Page 63 Manual No. M/97-164-F1 102Shear Stress (Dynes/Cm2)=TK * SMC * SRC * Torque= 0.09373 * 32.0 * 0.34 * 62.3

Page 61 - Appendix F - Speed Sets

Brookfield Engineering Labs., Inc. Page 64 Manual No. M/97-164-F1102Figure G1 Analog Output:The analog outputs for temperature and % torque are acc

Page 62 - Appendix G - Communications

Brookfield Engineering Labs., Inc. Page 65 Manual No. M/97-164-F1 102Appendix H - Model S Laboratory StandEXPLOSION-PROOFCLAMP VS-17Y PARTSBROOKFIE

Page 63

Brookfield Engineering Labs., Inc. Page 66 Manual No. M/97-164-F1102 UnpackingCheck carefully to see that all the components are received with no c

Page 64 - Analog Output:

Brookfield Engineering Labs., Inc. Page 67 Manual No. M/97-164-F1 102 Appendix I - DVE-50 Probe ClipProbe Clip DVE-50 is supplied with all model DV

Page 65 - Figure H1

Brookfield Engineering Labs., Inc. Page 68 Manual No. M/97-164-F1102 Appendix J - Fault Diagnosis and TroubleshootingListed are some of the more co

Page 66

Brookfield Engineering Labs., Inc. Page 69 Manual No. M/97-164-F1 102 ❏ Inaccurate Readings✓ Verify spindle, speed and model selection.✓ If % read

Page 67 - Figure H3

Brookfield Engineering Labs., Inc. Page 7 Manual No. M/97-164-F1 102I.5 Safety Symbols and PrecautionsSafety SymbolsThe following explains safety s

Page 68 - ❏ Spindle Does Not Rotate

Brookfield Engineering Labs., Inc. Page 70 Manual No. M/97-164-F1102If the above do not rectify the problem, do the following:✓ Shut off viscometer

Page 69 - ❏ Inaccurate Readings

Brookfield Engineering Labs., Inc. Page 71 Manual No. M/97-164-F1 102✓ Press ENTER/AUTORANGE. Information similar to Figure J4 will appear on your

Page 70 - READY PRINTER

Brookfield Engineering Labs., Inc. Page 72 Manual No. M/97-164-F1102 Appendix K - Warranty Repair and Service WarrantyBrookfield Viscometers are gu

Page 71 - TURN POWER OFF

Brookfield Engineering Labs., Inc. Page 73 Manual No. M/97-164-F1 102❏ Remove and return all spindles (properly packedfor shipping).❏ Clean excess

Page 72 - Warranty

Brookfield Engineering Labs., Inc. Page 74 Manual No. M/97-164-F1102Providing us with the following information will help us to service your equipm

Page 73 - Foam Insert

Brookfield Engineering Labs., Inc. Page 75 Manual No. M/97-164-F1 102VISCOSITY TEST REPORTSAMPLE MODEL SPINDLE RPMDIAL READINGFACTORVISCOSITY SHEAR

Page 74 - (please list all):

Brookfield Engineering Labs., Inc. Page 8 Manual No. M/97-164-F1102 SELECTDISPLAYSELECT DISPLAYSelects the data parameter to be displayed:cP Viscos

Page 75 - VISCOSITY TEST REPORT

Brookfield Engineering Labs., Inc. Page 9 Manual No. M/97-164-F1 102I.7 CleaningBe sure to remove the spindle from the instrument prior to cleaning