Brookfield Engineering Laboratories, Inc. Page 32 Manual No. M/85-150-P700
Inaccurate Readings
Verify spindle, speed and model selection
Verify test parameters: temperature, container, volume, method. Refer to:
"More Solutions to Sticky Problems"; Section II.2a — Considerations for Making
Measurements
Dial Viscometer Operating Manual; Appendix B — Viscosity Ranges
Dial Viscometer Operating Manual; Appendix C — Variables in Viscosity Measure-
ment
Perform a calibration check. Follow the instructions in Appendix D.
Verify tolerances are calculated correctly.
Verify calibration check procedures were followed exactly
If the unit is found to be out of tolerance, the unit may be in need of service. See Appendix G for
details on how to return your viscometer.