Brookeld Engineering Labs., Inc. Page 26 Manual No. M98-350-J0912
air bubbles beneath the disk by rst immersing the spindle at an angle, and then connecting
it to the viscometer.
4) The viscosity standard uid, together with the spindle and guard leg (if supplied), should
be immersed in the bath for a minimum of 1 hour, stirring the uid periodically, prior to
taking measurements.
5) After 1 hour, check the temperature of the viscosity standard uid with an accurate thermometer.
Fluid must be within ± 0.1°C of the specied temperature, normally 25°C. Allow longer
soak time if required to come to test temperature.
6) If the uid is at test temperature, measure the viscosity and record the viscometer reading.
Note: The spindle must rotate at least ve (5) times before readings are taken.
7) The viscosity reading should equal to the cP value on the viscosity uid standard to within
the combined accuracies of the viscometer and the standard (as discussed in Interpretation
of Calibration Test Results at the end of this Appendix).
Calibration Procedure for a Small Sample Adapter
When a Small Sample Adapter is used, the water jacket is connected to the water bath and the water
is stabilized at the proper temperature:
1) Put the proper amount of viscosity standard uid into the sample chamber. The amount varies
with each spindle/chamber combination. (Refer to the Small Sample Adapter instruction
manual.)
2) Place the sample chamber into the water jacket.
3) Put the spindle into the test uid and attach the extension link, coupling nut and free hanging
spindle (or directly attach the solid shaft spindle) to the DV-E.
4) Allow 30 minutes for the viscosity standard, sample chamber and spindle to reach test
temperature.
5) Measure the viscosity and record the viscometer reading.
Note: The spindle must rotate at least ve (5) times before a viscosity reading is taken.
Calibration Procedure for a Thermosel System
A two-step process is recommended for the Thermosel.
A) Evaluate the calibration of the Viscometer alone according to the procedure outlined in
this section, entitled Calibration Procedure for LV #1-3 (#61-63) and RV, HA, HB #2-6
Brookeld spindles.
B) Evaluate the Viscometer with the Thermosel according to the procedure described below.