Brookfield DV-I Prime Manuel d'utilisateur Page 63

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 71
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 62
Brookeld Engineering Laboratories, Inc. Page 63 Manual No. M/07-022-D0613
repeatable value can be obtained without any special effort for any measurement circumstance. But, it
should be known that this type of torque reading will not convert into a correct centipoise value when
using a Brookeld factor if the boundary conditions are not those specied by Brookeld.
The guard leg is a part of the calibration check of the Brookeld LV and RV series Viscometer/Rheometer.
Our customers should be aware of its existence, its purpose and the effect that it may have on data.
With this knowledge, the viscometer user may make modications to the recommended method of
operation to suit their needs.
Figure F-1
Vue de la page 62
1 2 ... 58 59 60 61 62 63 64 65 66 67 68 69 70 71

Commentaires sur ces manuels

Pas de commentaire