Brookfield Engineering Labs., Inc. page 35 Manual No. M14-223
To select the type of measuring system, press the appropriate radio button:
Cone
Plate
Vane
Cylinder
DG
Custom
By selecting “Visible” you determine whether users selecting a measuring system in the measuring
program can see this measuring system.
Press OK to acknowledge your choice. On the next screen you enter the shear rate factor (KGamma)
and the shear stress factor (Tau Promill) using the numeric keyboard and acknowledge your entries with
OK:
Figure III-26
For cones, truncation must be entered in "m on the next screen:
Figure III-27
Acknowledge your entry with OK. Then enter the sample volume required for the measurement in
milliliters (mL) in the next mask and acknowledge with OK.
Figure III-28
Now your new or changed measuring system is durably stored in your RST-CPS Rheometer.