Brookfield Engineering Labs., Inc. page 43 Manual No. M14-223
III.8.1 Time
Set the current date and the current time here and acknowledge with OK:
Figure III-49
III.8.2 Counter
This screen reports the number of tests that have been run by your RST Rheometer.
III.8.3 Zeroing
Zeroing measures the bearing friction and uses this information to adjust measurements during tests.
Figure III-50
High Range Calibration is zeroing for high speeds (1…1300 rpm) and Low Range Calibration is zeroing
for low speeds (0.01…1 rpm), and very low range calibration is zeroing for 0.01 rpm. If in doubt, run High
Range Calibration for a period of approx. 20 min.
III.8.4 RS232
This screen configures the RS232 port:
Figure III-51
III.8.5 Memory
This screen tells you how much free memory is available in your RST Rheometer. When you open the
“Folder” icon, you can review data files stored in memory.