SPECIALISTS IN THE
MEASUREMENT AND
CONTROL OF VISCOSITY
TEL 508-946-6200
F
AX 508-946-6262
or 800-628-8139 (USA excluding MA)
I
NTERNET
http://www .brookfieldengineering.com
BROOKFIELD ENGINEERING LABORATORIES, INC.
11 Commerce Boulevard, Middleboro, MA 02346 USA
with offices in
:
BostonChicagoLondonStuttgartGuangzhou
BROOKFIELD CT3
TEXTURE ANALYZER
Operating Instructions
Manual No. M08-372-C0113
Vue de la page 0
1 2 3 4 5 6 ... 55 56

Résumé du contenu

Page 1 - Manual No. M08-372-C0113

SPECIALISTS IN THEMEASUREMENT ANDCONTROL OF VISCOSITYTEL 508-946-6200FAX 508-946-6262or 800-628-8139 (USA excluding MA)INTERNET http:/

Page 2

Brookeld Engineering Labs., Inc. Page 10 Manual No. M08-372-C0113 RESET / STOP This key is used to stop a test in progress and return

Page 3 - Table of Contents

Brookeld Engineering Labs., Inc. Page 11 Manual No. M08-372-C0113to B Male. See Appendix D, (CT3 to Computer Command Set)2) RS-232 Serial Port

Page 4

Brookeld Engineering Labs., Inc. Page 12 Manual No. M08-372-C0113II. QUICK START 1. Unpack the instrument according to Section I.5. 2. Ins

Page 5 - I. INTRODUCTION

Brookeld Engineering Labs., Inc. Page 13 Manual No. M08-372-C0113III. OPERATIONIII.1 PrincipleTheBrookeldCT3TextureAnalyzercanbeopera

Page 6

Brookeld Engineering Labs., Inc. Page 14 Manual No. M08-372-C01131) Immediately stop probe movement and display the emergency stop screen.2)

Page 7

Brookeld Engineering Labs., Inc. Page 15 Manual No. M08-372-C0113itintoposition.Thetableisalsomountedonathreadedshaftheldsecurely

Page 8

Brookeld Engineering Labs., Inc. Page 16 Manual No. M08-372-C0113III.6 Operating MenuThe CT3 Texture Analyzer offers seven test modes, one ca

Page 9 - BROOKFIELD

Brookeld Engineering Labs., Inc. Page 17 Manual No. M08-372-C0113All parameters can be set within the ranges shown in Table III.1. Load Cell R

Page 10 - BACK PANEL

Brookeld Engineering Labs., Inc. Page 18 Manual No. M08-372-C0113III.7 Running A TestA test is initiated by depressing the start button. The

Page 11

Brookeld Engineering Labs., Inc. Page 19 Manual No. M08-372-C0113travelspeedis4.5mm/s.Duringthetesttherunningscreenisseenasshownb

Page 13 - III. OPERATION

Brookeld Engineering Labs., Inc. Page 20 Manual No. M08-372-C0113Deformation and Load will remain live displays during the test. When the count

Page 14 - Brookfield

Brookeld Engineering Labs., Inc. Page 21 Manual No. M08-372-C0113Testresultsareshownonthereportingscreen:CYCLE:XX PEAK MEAN:

Page 15

Brookeld Engineering Labs., Inc. Page 22 Manual No. M08-372-C0113NOTE: It is not recommended to conduct bloom testing with load cells greater

Page 16

Brookeld Engineering Labs., Inc. Page 23 Manual No. M08-372-C0113Springiness is a measure of how far the sample returns after being compressed

Page 17 - Load Cell Range

Brookeld Engineering Labs., Inc. Page 24 Manual No. M08-372-C0113Duringthetesttherunningscreenisseenasshownbelow:TEST: TENSIONDEFOR

Page 18

Brookeld Engineering Labs., Inc. Page 25 Manual No. M08-372-C0113withthesample.Oncetheprobecontactsthesamplethetestwillbeginatth

Page 19

Brookeld Engineering Labs., Inc. Page 26 Manual No. M08-372-C0113The return speed after the target is reached is the post test speed which is c

Page 20

Brookeld Engineering Labs., Inc. Page 27 Manual No. M08-372-C0113Hanger MountPrimary hanger to which additional mass is addedFigure III.342. T

Page 21

Brookeld Engineering Labs., Inc. Page 28 Manual No. M08-372-C0113Example of Calibration VericationA4500gramCT3TextureAnalyzer,according

Page 22

Brookeld Engineering Labs., Inc. Page 29 Manual No. M08-372-C0113IV. TEST METHOD DEVELOPMENTThe measurement results provided by the CT3 Textur

Page 23

Table of ContentsI. INTRODUCTION ...5I.1 Com

Page 24

Brookeld Engineering Labs., Inc. Page 30 Manual No. M08-372-C0113Withinaprobecategory,variationsofgeometrycanbesignicant.Itmayreq

Page 25 - TEMP. C/F: 19.0C

Brookeld Engineering Labs., Inc. Page 31 Manual No. M08-372-C0113V. APPLICATIONSThisapplicationssectionshouldbeusedasaguidetothedeve

Page 26

Brookeld Engineering Labs., Inc. Page 32 Manual No. M08-372-C0113 READING: Peak Positive ForceHARDNESSTotal Negative AreaAdhesiveness (Indica

Page 27

Brookeld Engineering Labs., Inc. Page 33 Manual No. M08-372-C0113ADHESIVEFORCE Force required to “pull” probe from sample(suction).MOD

Page 28

Brookeld Engineering Labs., Inc. Page 34 Manual No. M08-372-C0113 BIBLIOGRAPHY AND REFERENCES:Anon. 1988. Facts About Yogurt. National Dairy

Page 29 - IV. TEST METHOD DEVELOPMENT

Brookeld Engineering Labs., Inc. Page 35 Manual No. M08-372-C0113 READING: FHardness(H1)Quantity of FracturesCalculate mean of all values(HA

Page 30

Brookeld Engineering Labs., Inc. Page 36 Manual No. M08-372-C0113 CONCLUSIONS: Crispiness is an extremely complex parameter to quantify being

Page 31 - V. APPLICATIONS

Brookeld Engineering Labs., Inc. Page 37 Manual No. M08-372-C0113 READING: CONSISTENCYTotal +ve AreaHARDNESSPeak +ve ForceFADHESIONTotal -ve A

Page 32 - Fat Free

Brookeld Engineering Labs., Inc. Page 38 Manual No. M08-372-C0113A penetration depth of 4mm (approx. 50% deformation or strain) wasc

Page 33 - FACTORS: 1. Sample size

Brookeld Engineering Labs., Inc. Page 39 Manual No. M08-372-C0113Appendix A - Probes, Fixtures, Calibration Accessories and Gelatin Accessories

Page 34

Appendix A - Probes, Fixtures, Calibration Accessories and Gelatin Accessories ... 39 A-1 Probes ...

Page 35 - COLLECTION:

Brookeld Engineering Labs., Inc. Page 40 Manual No. M08-372-C0113 TA11/100 25.4mm AOAC STANDARD CYLINDER PROBE Clear Acrylic. 35mm Long. (Lo

Page 36

Brookeld Engineering Labs., Inc. Page 41 Manual No. M08-372-C0113TA15/100 45º CONE PROBEClear Acrylic. 30mm Diameter. (Low mass for CT3100) T

Page 37 - each penetration depth

Brookeld Engineering Labs., Inc. Page 42 Manual No. M08-372-C0113TA43 * 25.4mm DIAMETER BALL PROBENylon 14gTA49 25.4mm DIAMETER

Page 38

Brookeld Engineering Labs., Inc. Page 43 Manual No. M08-372-C0113 TA52 Small Scale MOHRS Shear BladeIncludesonebladeholderandveblades.

Page 39 - * included in TA-P-KIT2

Brookeld Engineering Labs., Inc. Page 44 Manual No. M08-372-C0113ACCESSORIES FOR FIXTURE BASE TABLETA-BT-KIT FIXTURE BASE TABLERectangular bas

Page 40 - * included in TA-P-KIT2

Brookeld Engineering Labs., Inc. Page 45 Manual No. M08-372-C0113TA-CTP COMPRESSION TOP PLATERectangular plate 150mm X 100mm, used for compres

Page 41

Brookeld Engineering Labs., Inc. Page 46 Manual No. M08-372-C0113TA-GPJ GENERAL PEELING JIGThisxturemeasurestheadhesivestrengthneededt

Page 42

Brookeld Engineering Labs., Inc. Page 47 Manual No. M08-372-C0113TA-MTP MAGNUS TAYLOR PROBE KITThis probe kit is used for puncture tests to me

Page 43

Brookeld Engineering Labs., Inc. Page 48 Manual No. M08-372-C0113TA-SBA SET OF 4 SHEAR BLADESIncluding Warner Bratzler test jig complete

Page 44

Brookeld Engineering Labs., Inc. Page 49 Manual No. M08-372-C0113TA-VBJ VOLODKEVITCH BITE JAWSCompression test used to simulate the shearing a

Page 45

Brookeld Engineering Labs., Inc. Page 5 Manual No. M08-372-C0113I. INTRODUCTIONThe CT3 Texture Analyzer is the most recent development of the o

Page 46

Brookeld Engineering Labs., Inc. Page 50 Manual No. M08-372-C0113TA-CW-10KGC CALIBRATION WEIGHT SET, 10kg WITH CASE & CERTIFICATESet of wei

Page 47

Brookeld Engineering Labs., Inc. Page 51 Manual No. M08-372-C0113Appendix B - TroubleshootingIf no power to the instrument:Check:• Userreplac

Page 48

Brookeld Engineering Labs., Inc. Page 52 Manual No. M08-372-C0113Appendix C - Texture Loader ProgrammingOVERVIEWTexture Loader software is prov

Page 49

Brookeld Engineering Labs., Inc. Page 53 Manual No. M08-372-C0113Userdenedtestwindow1)Testmode:Choosethetypeoftestyouwishtocreat

Page 50

Brookeld Engineering Labs., Inc. Page 54 Manual No. M08-372-C0113Appendix D - CT3 Connector Wiring Serial (RS-232) Communications Parameters

Page 51 - Appendix B - Troubleshooting

Brookeld Engineering Labs., Inc. Page 55 Manual No. M08-372-C0113Appendix E - Online Help and Additional Resourceswww.brookeldengineering.com*

Page 52 - OPERATION OF TEXTURE LOADER

Brookeld Engineering Labs., Inc. Page 56 Manual No. M08-372-C0113Appendix F - Warranty and Repair ServiceBrookeldTextureAnalyzersareguaran

Page 53

Brookeld Engineering Labs., Inc. Page 6 Manual No. M08-372-C0113I.2 UtilitiesInputVoltage: 90-265VACInputFrequency: 50/60HzPowerConsumpt

Page 54

Brookeld Engineering Labs., Inc. Page 7 Manual No. M08-372-C0113TriggerPoint: Range Resolution Model Grams Grams CT3-100 0.1-10 0.01 C

Page 55

Brookeld Engineering Labs., Inc. Page 8 Manual No. M08-372-C0113 4) Remove any additional components from the shipping package. Save the shi

Page 56

Brookeld Engineering Labs., Inc. Page 9 Manual No. M08-372-C0113 Functional Earth Terminal - Main power entry module must have an earth cond

Commentaires sur ces manuels

Pas de commentaire