Appendix J - Fault Diagnosis and Troubleshooting
Listed are some of the more common problems that you may encounter while using your Viscom-
eter.
Spindle Does Not Rotate
1. Incorrect power supply
Check - must match Viscometer requirements
2. Viscometer not plugged in
Connect to appropriate power supply
3. Power switch in “off” position
Turn power switch on
Spindle Rotates Eccentrically
1. Viscometer is not level
Adjust the laboratory stand
2. Spindle not screwed securely to coupling
Tighten
3. Dirt in spindle or Viscometer coupling
Clean
4. Bent spindle
Check other spindles - replace any that are bent
If all rotate eccentrically, contact Brookfield Engineering Laboratories, Inc. or your
Brookfield dealer for repair
Note: Maximum permissible runout is 1/16 inch (1.6 mm) at end of spindle.
5. Spindle coupling bent
Contact Brookfield Engineering Laboratories, Inc. or your Brookfield dealer for
repair
No Display Reading
1. Underrange “—” (in cP or SS) when % torque is below zero
Change spindle and/or speed
Perform an Auto Zero