Brookeld Engineering Labs., Inc. Page 26 Manual No. M13-167-B0614
Figure III-3
Figure III-4
Single Point:CollectonlyasingledatapointwhentheEndConditionismet.
Single Point Averaging:Specifyanamountoftimeoverwhichtoaveragemeasured
data.CollectasingledatapointwhentheEndConditionis
met.Thisdatapointisanaveragedvalue.Ifthetimefor
averagingisshorterthanthetotaltimeforthestep,thenthe
averagewillbeperformedforthespeciedtimeattheend
ofthetest.
Example1: TheEndConditionisTime withavalueof1 minuteand30seconds,
theSinglePointAveragingDurationis30seconds,thesingledatapoint
collectedfromthisstepwillbeanaverageofthedatameasuredfrom1
minute to 1 minute 30 seconds.