Brookeld Engineering Labs., Inc. Page 53 Manual No. M13-167-B0614
IV.4.5 Default Path
ThePathisthelocationwheredataortestsarestored.TheDV2Twillallowforstorage
totheInternalMemoryortoanattachedUSBFlashdrive.Thepathwillalsoincludeany
lestructurethathasbeencreated.
TheDefaultPathidentieswhatlocationwillbeusedastheinitiallocationwhenResultsare
saved. The Default Path dened in Reports - Congure ViscosityTest (saved test or
unsavedtest)willtakepriorityovertheDefaultPath.
TheDV2TcancreatelestructurebasedonUserIDandTestName.Theoptionsprovided
fortheDefaultPathspecifywhatformatthedatapathshouldutilize.Forexample:set
theDefaultPathtoOperator->TestName.TheUserisloggedinasDAVIDandthetest
torunisINK5.ThedatapathwillbedefaultedasshowninFigureIV-10.
Figure IV-10
TIP:TheDefaultPathcanbeoverwrittenatthetimethatdataissaved.
IV.4.6 Settings Reset
SettingsResetwillreturntheDV2Tsettings(Device,UserandGlobal)tothefactorydefault.
Internal memory will not be affected. User accounts will not be affected.
IV.4.7 Device Reset
DATA LOSS POSSIBLE!
DeviceResetwillreturntheDV2Ttothefactorydefaultcondition.Alldata,testsandAudit
Trailswillbedeletedfrominternalmemory.AllUseraccountswillbedeleted.
TIP:MovedatalesandtestlestoaUSBUSBFlashDrivepriortoselectingDevice
Reset.