CollectSinglePointatStepEnd.Duringthetest,thetotaltimerequiredto
reach200cPis10minutes40seconds.Totalpointscollectedwillbe11
withthelastdatapointtakenat10minutes40seconds,anaverageofthe
last20secondsofthestep.
No Data: EndConditionismetandnodataiscollected.
III.6 End Condition
ThecompletionofatestisdenedbytheEndCondition.EachtimethatyouenterCongure
ViscosityTest,theEndConditionwillbesettothelastvalueused.TheEndCondition
parameterorvaluescanbechangedbypressingtheEndConditionbutton.
Within the End Conditionscreen, the currentlyselected End Conditionparameter and
valuesaredisplayed. TheEnd Conditioncanbechanged by pressingthebluedown
arrowwithintheparametereld.SixEndConditionsareavailable:Time,#ofPoints,#of
Revolutions,Torque,Viscosity,Temperature(seeFigureIII-5).
Figure III-5
Time:Thetestwillcompletewhenthespeciedamountoftimehaselapsed.Timeis
enteredinHours,Minutes,andSeconds.
Hours: 0-99
Minutes: 0-59
Seconds: 0-59
Atimeofzerohours,minutesandsecondsmaybeselected.WiththisEnd
Condition,theDV2Twilloperateattheselectedspeeduntiltheoperatorselects
StopTest.DatawillbecollectedaccordingtotheDataCollectionsetting(see
SectionIII.5).
TIP:AnEndConditionofzeroTimecanbeusefulwhenmeasuringanew
material.Duringthetest,thespeedcanbechangedwithoutendingand