Vue de la page 40
Brookeld Engineering Labs., Inc. Page 41 Manual No. M13-167-B0614
AutoZero: AutoZero is an operation performed by the DV2T Viscometerautomatically
duringthepowerupsequence.Thisoperationsetstheinstrumenttransducer
toacorrectzerovalue.Thezerovalueshouldnotshiftovertime,however,
iftheuserdeterminesthatthezeroisnotcorrect,youcanforceanAutoZero
tobeperformedwithoutpoweringdowntheviscometer.WhentheAutoZero
functionisselected,theDV2TisimmediatelysettoAutoZeromodeandthe
userispresentedwiththeAutoZeroscreen(FigureII-2).Uponcompletion,the
userwillbereturnedtotheDeviceSetupmenu.
Oscillation Check: The Oscillation Check is a check of the performance of the lower
bearingoftheDV2TViscometer.ThislowerbearingiseitherPoint
andJewelorBallBearing.Correctperformanceofthelowerbearing
is essential for proper operation of the viscometer. A failed lower
bearingwillnormallycauseviscositymeasurementstobelowerthan
expected.
AnOscillationCheckisperformedasfollows:
• RemovethespindleandleveltheDV2TViscometer
• Gentlypushupontheviscometercoupling.
• Turnthecouplinguntilthe%Torquereadingis10–15%.
• Gentlyletgoofthecoupling.
• Watchthereading.The%Torquereadingshoulddecrease
smoothlyandoscillateabout0.0%.
SelectingOscillationCheckintheDeviceSetupmenuwillpresentthe
userrstwithascreeninstructingtoleveltheviscometerandremove
thespindle.Conrmationoflevelandspindleremovalwillpresentthe
userwiththeOscillationCheckscreenasshownbelow.
Figure IV-3
AsuccessfulOscillationCheckwillshowaverysmoothdecreaseof
%Torquewithanalvaluewithin+/-0.2of0.0.Anyvalueabove0.2%
orbelow-0.2%indicatesthatacalibrationcheckshouldbeperformed
(seeAppendixF).
TIP:HigherTorqueranges(HAandHB)willquicklyreturntothe
zerovaluewithoutdisplayingmuchoscillation.Thisisnormal
behavior.